Model |
KD-269 |
Measuring Range |
32.00¢XC ~ 42.99¢XC (89.60¢XF~109.39¢XF)
Below 32.00¢XC (89.60¢XF) displays L¢XC (¢XF)
Above 42.99¢XC (109.39¢XF) displays H¢XC (¢XF) |
Resolution |
0.01¢XF (0.01¢XC) |
Accuracy |
¡Ó0.10¢XC (34.00¢XC to 42.00¢XC); ¡Ó0.20¢XC (other range)
¡Ó0.20¢XF (93.20¢XF to 107.60¢XF); ¡Ó 0.40¢XF (other range)
|
Display |
Liquid Crystal Display 4 1/2 digits |
Battery (included) |
Micro Alkaline 192 or LR 41 (1.55V) |
Power Consumption |
0.15 milliwatt per measurement |
Battery Life |
More than 200 hours of continuous operation |
Dimension |
133.8 mm x 17.3 mm x 9.4 mm |
Weight |
Approx. 10.6 grams including battery |
Beeper |
Approx. 8 seconds sound signal when peak temperature reached |
Memory |
Built-in memory enabling display of the previously measured value |
Operating Conditions |
Temperature: 10 to 40¢XC ( 50 to 104¢XF)
Relative Humidity: 15 to 95 % |
Storage Conditions |
Temperature: -20 to 50¢XC (-4 to 120¢XF)
Relative Humidity: 15 to 95 % |
Guarantee of Quality |
Certification ISO13485
Conformity EN 12470-3, EN60601-1, and ASTM E1112-00 |
Low Battery Indication |
Replace the battery if the low battery indicator appears |
Auto Off |
Automatically after 10 minutes of non use |
Product classification |
Type BF equipment |